@misc{oai:u-fukui.repo.nii.ac.jp:00023294, author = {上田, 正紘 and 石川, 和彦 and 陳, 捷 and 當間, 安厚 and UEDA, Masahiro and ISHIKAWA, Kazuhiko and JIE, Chen and TOUMA, Yasunori}, month = {Oct}, note = {A method of high sensitive detection for measurement of blood leak has been proposed and demonstrated experimentally. The method was based on multireflection of light at a sidewall mirror of a container and normalization of an output light power to an input power. The obtainable maximum sensitivity for blood concentration was 0.1 ppm which was about 50 times higher than the value of the existing sensors.}, title = {高感度光漏血センサー}, year = {1993} }